Recalled Apr 30, 2026 · #26449

Adult Portable Bed Rails Recalled Due to Risk of Serious Injury or Death from Entrapment and Asphyxiation; Violate Mandatory Standard for Adult Portable Bed Rails; Imported by Hepo Care Medical Equipment Online

Hazard

The recalled bed rails violate the mandatory standard for adult portable bed rails because users can become entrapped within the bed rail or between the bed rail and the side of the mattress, posing a serious entrapment hazard and risk of death by asphyxiation. The bed rails also do not meet structural stability or retention strap requirements, posing a fall hazard. The bed rails' push pins and push pin holes are incorrectly sized, posing a laceration hazard. In addition, the bed rails do not bear the required hazard warning labels.

Be the first to hear about future recalls.

Add the things you own to a watchlist. We’ll alert you the moment one matches.

Get started, free

Description

This recall involves Hepo-branded adult portable bed rails, model 110039. The recalled bed rails have white tubing with black foam rubber handle grips, a black fabric pouch, and measures about 16 inches wide by 25 inches high. The model number is located on the original product packaging.

Product details

Product
Hepo Adult Portable Bed Rails
Type
Manufacturer
Mobility Source Medical Technology Co., Ltd. of China
Model numbers
110039
UPCs
Sold at
Online at Walmart.com from November 2025 through January 2026 for about $100 Amazon.com from June 2025 through January 2026 for between $35 and $40.
Made in
China

What you should do

Consumers should stop using the recalled adult portable bed rails immediately and contact Hepo Care Medical Equipment Online for a full refund. Consumers will be asked to destroy the bed rails by dismantling the bed rails, cutting off the storage bag and writing "RECALLED" on the upper and lower rails with permanent marker, take a photo of the destroyed rails and upload the photo to www.hepolife.com/bed-rail-safety-recall or email the photo to manager@lqxmedical.com.

View the official CPSC notice